Product Details:
|
Usage: | Waterproof Universal Rain Spraying IP Test Equipment IPX5 / IPX6 | Rotating Speed: | Adjustable |
---|---|---|---|
Arm Radius: | 300mm, 400mm, 600mm, 800mm, 1000mm, 1200mm Or 1500mm | Rotating Angle: | +/-270, Adjustable |
Rotating Table: | Speed Is Adjustable | Flow Rate: | 0.07L/min (IPX3,4), 0.6+/-0.5L/min (IPX4K) |
Nozzle Distance: | 50mm | Nozzle Diameter: | 0.4mm |
Power: | 1 Phase, 220V, 5KW | Specimen Rotative Speed: | 1-3r/min (Adjustable) |
Highlight: | ip testing equipment,ingress protection testing |
IPX3 / IPX4 Programmable IP Testing Equipment Computer Controller Rain Spray Test Room
Haida IPX1 - IPX8 Rain Test Chamber.pdf
Description:
IPX1-IPX9 Programmable IP Test Equipment Computer Controller Rain spray Test Room is suitable for electric products assessment, shell and seal in the rain environment can ensure that equipment and components good performance test. This product uses scientific design, makes the equipment can realistic simulation drop of water, water spray, splash water, water spray, etc., various kinds of environment. Into the comprehensive control system and the adoption of the frequency conversion technology, which makes the rainfall test product frame rotation Angle, jet pendulum rod swing Angle of water quantity and the oscillating frequency can be automatic control.
Auxiliary Structure
Bottom of the chamber has a water storage tank, test water sprinkler systems, table rotation system, swing pipe swing drive
Seal: double high-temperature high tensile seal between the door and the cabinet in order to ensure a closed test area
Door handle: no reaction door handle, easier operation
Casters: bottom of the machine can be fixed with high quality PU wheels
Control System
1, a computer system using win 7
3, has a history memory function (available historical records test within 7 days)
4, temperature: 0.1 ℃ (display range)
5, Time: 0.1min
Product Usage
Rain chamber is mainly applied to electric and electronic, aerospace, military and other scientific research units, external lighting, automotive lighting and signaling devices test for detection of shell protection.
Box Structure
Tank shell material made of high quality stainless steel hairline, liner material stainless steel light board; 2 large sight glass door for easy observation test cabinets test sample status;
Imported inverter speed control to ensure the test according to the standard;
Bottom of the chamber can be fixed with high quality PU wheels, easy to move users;
It has a 270 degree swing pipe and 360-degree rotating rod Sprinklers
Adjustable speed of the sample stage
Jet Nozzle (Meet IPX5 and IPX6 test)
1. 6.3mm nozzle diameter, for IPX5 test. Water flow: 12.5L/min.
2. 12.5mm nozzle diameter, for IPX6 test. Water flow: 100L/min.
3. Meet IEC60529, IEC60335
4. Water Pumping System as option
Main Parameters of IPX1-IPX9 Programmable IP Test Equipment Computer Controller Rain spray Test Room
Standard Features
Item | Specification |
Internal dimension | (W*D*H)1000*1000*1000mm |
External dimension | (W*D*H)1300*1100*1700mm |
Oscillating tube | Radius 400mm |
Oscillating angle |
IPX3: spray ±60° from vertical line to each side, total 120° IPX4: spray ±180° from vertical line to each side, total 360° |
Oscillating speed |
IPX3: 2*120° for 4s IPX4: 2*360° for 12s |
Spray nozzle |
Diameter: 0.4mm, removable Ensure the water sprays in a line and without tolerance |
Distance between two nozzles | 50mm |
Turntable dimension | 500mm |
Turntable rotation speed | 1-3 cycles/min |
Water flow of each nozzle | 0.07 L/min +5% |
Test duration | 1~999,999min, adjustable |
Pressure gauge | 80-150Kpa |
Material | Chamber - Stainless steel ; Spray nozzle - Brass |
Features :
1. Rotating rain and spray nozzles designed to IPX specifications
2. Speed control for the rotating spray nozzles
3. Stationary product shelf - A rotating shelf is optional
4. Water pressure regulators, gauges and flow meters
5. Water circulation system to reduce water consumption
6. Adjustable swivel angle
7. Replaceable swivel tubes
8. Nozzle fittings can be swiveled
9. Exchangeable nozzle fittings
10. Adjustable water volume flow
11. Measurement of water volume flow
Operating Specifications
1, after the power is turned on when the machine setting control program has finished running, the machine will stop running;
2, when the control program is set to run is completed, the machine will stop running;
3, the door handle to open the box, put the sample into the test sample holder; then close the door;
Note: placing the sample volume must not exceed 2/3 of the capacity of the test area;
4. The "TEMI880 Operating Manual", the first test set operation, and then into the test state according to the set operating mode;
5, when observed in the test chamber Ruoyu situation changes, can open the door light switch, via Windows knows the situation changes inside the open; displays temperature and humidity test chamber on the controller (if no humidity test the humidity value without display);
6, open the box door handles, the test specimens were removed from the sample holder to view the sample after the test and record test condition; the test is completed;
7. After the test is completed, turn off the power switch.
Precautions
1, in operation accidentally hear the sound, the need to stop to check, to be isolated after troubleshooting before reboot, so as not to affect the service life of equipment.
2, the drive mechanism must be regularly refuel, reducer must be added # 20 clean oil.
3, after positioning the device, you need the support frame propped up against the device after the test casters subject to vibration displacement.
4, rain chamber for a long time to run, such as found in water clogging the pipeline must be removed, rinsed with tap water and then up the assembly.
Rain spray Test Room View:
(Rain Spray System)
Contact Person: Mary
Fax: 86-0769-89280809