Product Details:
|
Humiture Range: | 30-98%RH | Rainfall Cycle: | 1-240min |
---|---|---|---|
Certificate: | ISO9001: 2008 | Warranty Period: | 12 Months |
Studio Size: | 95Dx95Wx90Hcm | Box Material: | Stainless Steel SUS304# |
Insulation Materials: | Superfine Glass Fabric | Control System: | Germany KLOCKNER-MOELLER Progammable Controler |
Gas Processing: | Dischager For Fector And Harmful Gas | Use Environment: | 5°C~+30°C, ≤85%R.H |
High Light: | laboratory testing equipment,lab testing equipment |
Xenon Lamps Accelerated Aging Test Chamber Full Spectrum Xenon Aging Chamber
Product Introduction:
Sunlight, heat and moisture cause millions of dollars in product damage every year. Cracking, crazing, hazing, fading and yellowing can occur indoors or outdoors. With Accelerated Aging Chamber Machines, you can simulate the damage caused by full-spectrum sunlight, temperature and moisture. In just a few days or weeks, xenon lamps Aging Test Chamber can reproduce the damage that occurs over months or years outdoors
Useage
It can test the accelerated aging of rubber, plastic, fabric, textile, metal and so on.
Key Features
Remarkably Simple User Interface
AUTOCAL Calibration
Full Spectrum Xenon Lamps
Long-Life Optical Filters
Programmable Water Spray
SOLAR EYE Irradiance Control
Precision Temperature Control
Versatile Specimen Mounting
Relative Humidity Control
Auxiliary Structure
Seal: double high-temperature high tensile seal between the door and the cabinet in order to ensure that the test area
Door handle: no reaction door handle, easier to operate
Moves the sample holder: Sample holder rely installation box at the bottom of the linear reciprocating drive system for moving up and down
Casters: bottom of the machine can be fixed with high quality PU wheels
Main Parameter:
Item | Specification |
Internal Dimension (W*D*H) | 800*800*800mm |
External Dimension (W*D*H) | 1450*1500*2000mm |
Inner chamber material | SUS 304 # Stainless Steel, Anti-rust, easy to clean |
Outer chamber material | Stainless steel |
Chamber Temperature | 40±3℃ |
Temperature Uniformity | ≤3℃ |
Blank panel temperature | 65±3℃ |
Humidity Range | 50%-90% |
Irradiated intensity | 0.35±0.02w/m²at 340nm |
Xenon lamp power | 6.5KW water-cooled xenon lamp x 1pc |
Control system | PLC controller, touch screen |
Distance from specimen to arc center | 250~300mm, Effective exposure area 6400cm2 |
Max. Sample dimension | 200*90mm Thickness<20mm |
Filter | Daylight filter |
Door seal | Silicone rubber seal |
Temperature sensor | PT100 high precision temperature sensor |
Air circulation system | Stainless steel centrifugal rotor vane to improve the uniformity of temperature and humidity |
Cooling ventilator | Automatic controlled by temperature controller, and inner chamber cooled by refrigerating system |
Cooling ventilator | France Tecumseh compressor |
Power | AC380V, 50Hz, three phase |
Safety Protection |
Over heat protection for fan, Over load protection for cooling system, Over pressure protection for cooling system, Over temperature protection, Over heat and over flow protection for water pump, Leakage protection, Water shortage protection, Auto stop after failure warning |
Simple to Maintain
We believe that just because a product is technical, it doesn’t have to be hard to understand or difficult to maintain and repair. Instead of complicating our tester designs by loading them with extra or unnecessary features, we put our engineering effort into keeping things simple. Subsystems are modular, easy to troubleshoot and even easier to replace.
Operation
Xenon Aging Chamber are extremely simple to operate. Specimen mounting and evaluations are simplified with specially designed specimen holders. Programming is intuitive. All models are completely automated and can operate continuously, 24 hours per day, 7 days per week.
Simple to Afford
Accelerated Aging Chamber were specifically designed to have the lowest total cost of ownership in the industry. Their low purchase price, low lamp price, and low operating costs set a new standard for lightfastness testing. Now even the smallest lab can afford xenon arc weathering and lightstability testing.
Maintenance
1, any time to maintain a neat appearance;
2. Do not place objects on top of the machine to heat parts of Lima Da and control;
3, xenon lamp aging chamber after long-term use, spare parts, motor blower unit, wheel deceleration devices, if abnormal noise, should be checked, and pay attention to regular cleaning fuel;
4, electrical components such as over-temperature phenomena, must be checked immediately causes;
5, after each use, a xenon lamp aging test chamber should be kept clean to prevent corrosion;
6, the control cabinet clean-up once a month to prevent excessive dust, affecting the normal operation of equipment;
7. After the test is complete, you should turn off the main power, water stains wipe clean indoor work, try to keep the chamber in a dry environment;
8, the electrical components on the control panel, such as the need to exchange failure occurs, Do not indiscriminate demolition ensure xenon lamp aging chamber life.
Pictures:
Contact Person: Kelly
Fax: 86-0769-89280809