Product Details:
|
PH: | NSS ACSS6.5~7.2 CASS 3.0~3.2 | Power: | 1A C220V |
---|---|---|---|
Certificate: | ISO9001:2008 | Design Criteria: | CNS, JIS, K5400, Z2371, ASTM |
Spray Quantity: | 1.0~2.0 Ml / 80cm2 / Hr | Warranty Period: | 12 Months |
Brine Temperature: | 35°C±1°C | Test Chamber Temperature: | NSS ACSS 35±1°C CASS:50±1°C |
Temperature Precision: | ±1°C | Color: | White |
Highlight: | textile testing machine,textile testing instruments |
CE Approval SO2 Textile Weathering Testing Salt Spray Corrosion Test Chamber
Product image & Description:
CE Approval SO2 Salt Spray Corrosion Test Chamber can determine the corrosion resistance of iron metal or the iron metal inorganic film or organic film test, such as: electroplating, anode processing, conversion coating, painting, ect.
CE Approval SO2 Salt Spray Corrosion Test Chamber is an accelerated corrosion test that produces a corrosive attack to the coated samples in order to predict its suitability in use as a protective finish. The appearance of corrosion products (oxides) is evaluated after a period of time. Test duration depends on the corrosion resistance of the coating; the more corrosion resistant the coating is, the longer the period in testing without showing signs of corrosion. Different coatings have different behaviour in salt spray test and consequently, test duration will differ from one type of coating to another.
Characteristics:
1. spray nozzle the heart of teser adopt Bo Nute Principle which change saline water into atomization but no crystallization produced.
2.controller is digital type microcomputer which delivers precision temparature
3. parameters show on the display screen such as temparature, humidity, pressure, spray capacity and time.
4. test specifications are accept worldwild.
Product parameters:
Model | HD-E808-60 |
Internal Size (L×W×H) | 600x400x450mm |
External Size (L×W×H) | 1070x600×1180mm |
Equipment Material | Both inside & outside shell adopt P.V.C rigid plastic board |
Temperature Range | Room temperature ~ 55℃ |
Temperature Fluctuations | ≤±0.5℃ |
Temperature Uniformity | ≤±2℃ |
Temperature Precision | ±1℃ |
Spraying method | Continuous and Periodic spray type |
Test Chamber Temperature | Salt Spray Method (NSS ACSS)35±1℃ Corrosion-resistant Testing Method (CASS) 50±1℃ |
Saturated air Barrel Temperature | Salt Spray Method (NSS ACSS)47±1℃ Corrosion-resistant Testing Method (CASS) 63±1℃ |
Brine Temperature | 35℃±1℃ |
Spray Quantity | 1.0~2.0 ml / 80cm2 / hr |
PH value | Salt Spray Method (NSS ACSS6.5~7.2) Corrosion-resistant Testing Method(CASS)3.0~3.2 |
Lab Volume | 108L |
Brine Tank Capacity | 25L |
Multiple Safety Protection Devices | Current discharge protection, over pressure protection, over temperature protection, over load fuse protection |
Accessories | Testing Salt/ Measuring tank/Nozzle Equipment x 1 set |
Air source | 1HP Air pump (provided by buyer) |
Control system:
1, Temperature and humidity controller: imported Korea intelligent controller
2, Sensor: PT100 test sensor
3, LED digital display
4, Full digital multiple warning protection system, anti-dry and anti-over temperature
5, With SSR output, imported relay contact
6, Multiple time mode, S, M, H can freely switch
7, Multiple control method for Salt and SO2 switch
8, With defog and desulphuration function to ensure salt and SO2 avoid leakage
9, With flow meter to control the SO2 concentration by rate of flow and time
Chamber material:
1, Chamber shell: 8mm corrosion resistance PVC plate, 85℃high temperature resistance.
2, Sample holder: fiberglass rod, corrosion resistance.
3, Pressure barrel: 2.0mm SUS#304, seamless joint.
4, Heater: titanium alloy electric heater, corrosion resistance.
5, Machine support: SECC iron with paint spray, have wheels for sliding.
6, Chamber cover: with imported 8mm PVC acrylic transparent observation window.
7, Cover open method: SNS pneumatic method for uniform performance.
8, Machine cover and body are sealed with water to avoid gas leakage.
9, Gas pipe uses corrosion resistance silicone tube.
10, Equip with double gas filtration device to ensure the discharge gas is not harmful for environment.
Contact Person: Mary